Publication

A new parallel closing mechanism for the laminectomy rongeur makes it significantly more precise: a biomechanical and mechanical comparison study

Journal Paper/Review - Oct 24, 2017

Units
PubMed
Doi

Citation
Külling F, Ebneter L, Rempfler G, Zdravkovic V. A new parallel closing mechanism for the laminectomy rongeur makes it significantly more precise: a biomechanical and mechanical comparison study. Eur Spine J 2017
Type
Journal Paper/Review (English)
Journal
Eur Spine J 2017
Publication Date
Oct 24, 2017
Issn Electronic
1432-0932
Brief description/objective

PURPOSE
To prove that a modified closing mechanism of the rongeur gives better precision compared to the old Kerrison rongeur.

METHODS
Forty persons from the departments of orthopaedic surgery, urology and neurosurgery (35 orthopaedic, 2 urology and 3 neurosurgery) took part in the study. All participants were asked to punch ten times in a first step with either the old Kerrison rongeur with the scissors-like handle or the modified punch with a new parallel closing mechanism. In a second step, they punched 10 times with the other instrument. Shaft movement in three dimensions was measured with a stereoscopic, contactless, full-field digital image correlation system.

RESULTS
The new rongeur is significantly more precise with less movement in all three dimensions. The mechanical model of the new rongeur shows that the momentum needed to keep the tip at the initial position changes only minimally during the closing act on the new model.

CONCLUSION
The new rongeur is more precise compared to the old Kerrison model. It is more robust against changes in the direction of the finger forces and may reduce soreness, fatigue and CTS in spine surgeons.

LEVEL OF EVIDENCE
Not applicable: technical study.