Kantonsspital St.Gallen
login

Eleven-year results of the anatomic coated CFP stem in primary total hip arthroplasty

Daniel O Kendoff, Mustafa Citak, Claus Christoph Egidy, Padhraig F O'Loughlin & Thorsten Gehrke

abstract There is a paucity of intermediate term results relating to short stem prostheses. The current study represents the longest follow-up results of the CFP prosthesis to the authors' best knowledge. Between January 1999 and December 2000, all total hip arthroplasty patients (n = 149), treated with this anatomic neck preserving stem in the authors' institution were enrolled in this study. After a mean follow-up time period of 11.2 years 117 patients were available for the follow-up examination. The mean HHS increased from 53 to 93. Overall, revision surgery was required in 11 patients (9.4%), with implant-associated complications noted in only five cases (4.3%). Aseptic loosening leading to implant removal was noted in four patients (3.4%). The current study's data suggest that the CFP implant, used in young patients undergoing primary total hip arthroplasty, is safe and provides excellent results after 11 years.
   
citation Kendoff D O, Citak M, Egidy C C, O'Loughlin P F, Gehrke T. Eleven-year results of the anatomic coated CFP stem in primary total hip arthroplasty. J Arthroplasty 2013; 28:1047-51.
   
type journal paper/review (English)
date of publishing 20-03-2013
journal title J Arthroplasty (28/6)
ISSN electronic 1532-8406
pages 1047-51
PubMed 23523502
DOI 10.1016/j.arth.2012.10.013