Can plate osteosynthesis of periprosthethic femoral fractures cause cement mantle failure around a stable hip stem? A biomechanical analysis

Karlmeinrad Giesinger, Lukas Ebneter, Robert E Day, Karl K Stoffel, Piers J Yates & Markus S Kuster

abstract Periprosthetic femoral fractures (PFF) are a serious complication after total hip arthroplasty. Plate fixation with screws perforating the cement mantle is a common treatment option. The study objective was to investigate hip stem stability and cement mantle integrity under dynamic loading. A cemented hip stem was implanted in 17 composite femur models. Nine bone models were osteotomised just distal to the stem and fixed with a polyaxial locking plate the other eight constructs served as the control group. All specimens were tested in a bi-axial material testing machine (100000 cycles). There were no statistically significant differences in axial nor in medial (varus) stem migration. No cement cracks were detected in both groups. Plate fixation of a PFF with a stable, cemented prosthesis did not lead to cement mantle failure in this in vitro study.
   
citation Giesinger K, Ebneter L, Day R E, Stoffel K K, Yates P J, Kuster M S. Can plate osteosynthesis of periprosthethic femoral fractures cause cement mantle failure around a stable hip stem? A biomechanical analysis. J Arthroplasty 2014; 29:1308-12.
   
type journal paper/review (English)
date of publishing 3-1-2014
journal title J Arthroplasty (29/6)
ISSN electronic 1532-8406
pages 1308-12
PubMed 24439999
DOI 10.1016/j.arth.2013.12.015
contact Karlmeinrad Giesinger